== Re-flashing ==
== Re-flashing ==


The oscilloscope has required re-flashing several times. After pressing standby, If it needs re-flashing with new firmware, the LCD will light up, but the unit will not boot after pressing the front panel standby button.
The oscilloscope has required re-flashing several times. If after pressing the standby button the LCD lights up but does not boot after you waited for more than 20 seconds, then its likely the oscilloscope has experienced this issue again.  


Procedure:
Re-Flashing firmware procedure:
* Requires a windows machine. Works on XP, not tested on other machines.  
* Requires a windows machine. Works on XP, not tested on other machines.  
* USB-to-serial adapter.
* USB-to-serial adapter.
Line 49: Line 49:
* Connect the USB-to-serial adapter to the serial crossover cable
* Connect the USB-to-serial adapter to the serial crossover cable
* Connect the serial cable to the oscilloscope
* Connect the serial cable to the oscilloscope
* connect the USB-to-serial adapter to the PC
* Connect the USB-to-serial adapter to the PC. You may have to wait for drivers to install.
* Find which com port is being used in the windows control panel | system | hardware | Lpt and com option
* Find which com port is being used in the windows control panel | system | hardware | Lpt and com option
* Download firmware for the 820S from the [http://www.gwinstek.com/en-global/products/Discontinued_Products/Discontinued_Oscilloscope/GDS-800 Instek website]. This is located in the Discontinued section, under the download tab.  
* Download firmware for the 820S from the [http://www.gwinstek.com/en-global/products/Discontinued_Products/Discontinued_Oscilloscope/GDS-800 Instek website]. This is located in the 'Discontinued' section, under the GDS-800 series download tab. Ensure you download the correct software. There will be a version for other oscilloscopes in the same area.
* Unzip the instruction pdf file and exe program which performs the update.
* Unzip the instruction pdf file and exe program which performs the update.


At this point you can either follow the Instek Manual or these instructions.  
At this point you can either follow the Instek manual or these instructions.  
* Ensure the oscilloscope is turned off
* Ensure the oscilloscope is turned off
* Run the .exe
* Run the .exe
Line 64: Line 64:
* If not a timeout will occur and you will be asked to power off the oscilloscope and you will have to retry.
* If not a timeout will occur and you will be asked to power off the oscilloscope and you will have to retry.
* This may take several attempts before a download commences.
* This may take several attempts before a download commences.
* When download is completed, you will be asked to click next to program the software. Click next.  
* When download is completed, you will be asked to click next to program the software. Click next (doing a little jig is encouraged at this point).  
* After about 10 seconds you should then see the normal oscilloscope display.
* After about 10 seconds you should then see the normal oscilloscope display
* Press Exit on the PC programming software
* When the oscilloscope displays the normal oscilloscope screen. Press 'Exit' on the PC programming software
* Test that the oscilloscope reboots by taking the oscilloscope into and out of standby a few times.
* Test that the oscilloscope reboots by taking the oscilloscope into and out of standby a few times  


== Calibration ==
== Calibration ==

Revision as of 20:50, 2 April 2016

Oscilloscope
File:InstekGDS-820S.jpg
Digital oscilloscope
Equipment Information
Status Unknown
Manufacturer GW Instek
Model GDS-820S
Hackspace Information
Induction Required Yes
Induction Information Here
Card RFID Enabled No
Owner Hackspace & Members


The Basics

An oscilloscope is a piece of electrical test equipment which allows you to visually see an electrical signal shown on a display.

Specification

  • GW Instek GDS-820S
  • Digital storage oscilloscope (DSO)
  • 2 channel
  • Real Time sample rate: 100MS/s maximum on each channel
  • Equivalent Time sample rate: 25GS/s E.T. maximum on each channel
  • Bandwidth: 150MHz effectively an equivalent time bandwidth. So not strictly speaking an 150MHz scope unless you are talking about the equivalent time sample rate on a repetitive signal.
  • 5.7 inch mono LCD (320x240)
  • FFT, 125k/CH
  • RS232, USB and GPIB
  • Trigger DC-30MHz for 0.5div or 5mV, 30MHz to 150MHz 1.5 div or 15mV
  • Issues: Boots now, but waveform trigger doesn't appear to be working correctly. waveforms continuously roll.

Previous issues: Power board was repaired previously. The firmware was reflashed on two seperate occasions.

  • Manual: TBD
  • Owner: Fligg

Power on

Switch power on at the rear of the oscilloscope. Press the standby button in the top right of the front panel. Note that the button sometimes sticks.

Re-flashing

The oscilloscope has required re-flashing several times. If after pressing the standby button the LCD lights up but does not boot after you waited for more than 20 seconds, then its likely the oscilloscope has experienced this issue again.

Re-Flashing firmware procedure:

  • Requires a windows machine. Works on XP, not tested on other machines.
  • USB-to-serial adapter.
  • Serial 9 pin cross-over cable (e.g. the blue serial cable with 2 x female 9 pin D types)
  • Turn off the oscilloscope at the rear panel switch
  • Connect the USB-to-serial adapter to the serial crossover cable
  • Connect the serial cable to the oscilloscope
  • Connect the USB-to-serial adapter to the PC. You may have to wait for drivers to install.
  • Find which com port is being used in the windows control panel | system | hardware | Lpt and com option
  • Download firmware for the 820S from the Instek website. This is located in the 'Discontinued' section, under the GDS-800 series download tab. Ensure you download the correct software. There will be a version for other oscilloscopes in the same area.
  • Unzip the instruction pdf file and exe program which performs the update.

At this point you can either follow the Instek manual or these instructions.

  • Ensure the oscilloscope is turned off
  • Run the .exe
  • Choose which com port you are using
  • Click 'YES' to start the software
  • Turn on the oscilloscope at the rear and also press the front panel standby button.
  • You will need to do this in quick succession. To avoid a timeout.
  • At this point a download should commence.
  • If not a timeout will occur and you will be asked to power off the oscilloscope and you will have to retry.
  • This may take several attempts before a download commences.
  • When download is completed, you will be asked to click next to program the software. Click next (doing a little jig is encouraged at this point).
  • After about 10 seconds you should then see the normal oscilloscope display
  • When the oscilloscope displays the normal oscilloscope screen. Press 'Exit' on the PC programming software
  • Test that the oscilloscope reboots by taking the oscilloscope into and out of standby a few times

Calibration

The spaces oscilloscopes will NOT normally have a calibration certificate. Most will be well outside their calibration date. Having said that it is common to perform a basic oscilloscopes calibration check before each use against a known calibration source to confirm it is suitable for use. Most oscilloscopes have a calibration signal output point on the front panel. Some have rear panel outputs.

Induction

Ask on the mailing list for volunteers to demonstrate the use of any of the test equipment.

Safety

Unless you are very experienced with an oscilloscope then it is not advisable to probe high voltages or high power sources. It is possible to electrocute yourself or damage the oscilloscope. Special oscilloscope probes and specific connection methods are needed for probing high voltage and mains electricity signals. The oscilloscopes are replaceable, you are not! We want you to be safe.

Oscilloscopes can be heavy instruments. Bend at the knees when lifting and ask for help if you find the instrument too heavy to move.

Do not leave oscilloscope probes (or any wires) hanging over the edge of a bench or on the floor. You will stand on or trip over them.